Documenten - Waarschuwingen van fabrikanten medische hulpmiddelen
763 publicaties over Waarschuwingen van fabrikanten medische hulpmiddelen
- Datum
- Relevantie
-
Leonhard Lang GmbH RKL-4453 (GS ref.: VGL-15814) - GS defibrillation electrodes
In the course of an investigation triggered by a complaint, it was discovered that there is a possible contact problem between ...
-
Homoth Medizinelektronik GmbH & Co.KG FSCA-LK-HO-15112019 - Luftkalorisator, Air Calorisator
As a part of our repair evaluation, Homoth Medizinelektronik has detected that, due to a possible production error, the medical ...
-
Philips, FSCA 86100208A-IR, HeartStart XL Defibrillator/Monitor
Philips heeft een probleem geconstateerd bij de draaischakelaar voor therapieselectie van de HeartStart XL+ ...
-
Philips Goldway (Shenzhen) Industrial Inc FSCA 86100207A-IR Efficia DFM 100
De draaischakelaar voor therapieselectie van de Philips Efficia DFM100 Defibrillator/monitor kan uitvallen, waardoor het apparaat ...
-
Philips Goldway (Shenzhen) Industrial Inc. FSCA 86000260A-IR Efficia DFM 100
De Philips Efficia Defibrillator/monitor schakelt niet in of probeert onverwacht opnieuw op te starten, waardoor de Philips ...
-
Torunskie Zaklady Materialow Opatrunkowych, RI/22.08.2019, Matoset
One of our suppliers of the cardboard containers has found cases of non sterile containers in sterilised sets which could have ...
-
Angiokard Medizintechnik - sets with cardboard kidney dishes
With this letter we would like to inform you about a possible problem with the sterilisability of the cardboard kidney dish which ...
-
FSN Vitalhealth Engage - Philips VitalHealth Software B.V.
Philips VitalHealth Engage 5.1.0.4 vervangt de Hartfalen applicatie 1.0, die werd gebruikt door drie klanten in Nederland. ...
-
Eurovema Mobility AB 6.6.2-2019-47648 Euroflex work chairs with Forma chassis
Eurovema Mobility has discovered, that the screw that holds the chassis halves together on the older work chairs, can loose with ...
-
Varian Medical Systems CP-2019-01813 - Stainless Steel Interstitial Needles, 17 Gauge
Varian has received several reports that the trocar tip detached from a 17-gauge stainless steel interstitial needle.