Ortho Clinical Diagnostics Inc, CL2022-209, ORTHO Optix Reader ORTHO BioVue Cassettes

On the ORTHO Optix Reader, when reviewing the results, the user can edit a grade and/or analysis of the results. When performing this action, the user may also choose to cancel the edit. While cancelling the edit, the user is incorrectly informed that the ‘Changes have not been saved’ via a user interface prompt, which the user must acknowledge/accept. After the user accepts cancel of the edit, the result displayed on the user interface reverts to the original result and awaits result approval, as expected. When the user completes result approval (displayed with the original result), the result that is actually stored in the database and reported out of the system (such as to a connected LIS) is the edited result that the user thinks was previously cancelled.

Fabrikant: Ortho Clinical Diagnostics Inc
Product: ORTHO Optix Reader ORTHO BioVue Cassettes
Datum FSN: (documentdatum) 17-08-2022

Waarschuwing van de fabrikant

Dit bericht is een waarschuwing van de fabrikant. Na het op de markt brengen van een medisch hulpmiddel is de fabrikant verplicht het hulpmiddel te volgen en waar nodig te verbeteren. Wanneer sprake is van een risico voor de gebruiker moet de fabrikant actie ondernemen. De fabrikant informeert gebruikers en de inspectie over deze actie met een waarschuwing (Field Safety Notice). Zie ook waarschuwingen medische hulpmiddelen. Dit hoort bij het toezicht op medische technologie.