Spiegelberg GmbH and Co.KG, CMP-202211001

All the above-mentioned products can be tunneled away from the operation site, with the help of a trocar. After tunnelling, the air lumen of the probe is connected to an extension tube that in turn is connected to the Spiegelberg-ICP-Monitor.
To connect the extension tube, a securing nut is pushed on the probeside-connector. The securing nut contains two plates that allow the probeside-connector to snap securely in place inside the securing nut. Due to a slowly progressed joining-tool wear, it is possible that the two plates inside the securing nut might break free of the securing nut

Fabrikant: Spiegelberg GmbH & Co.KG,
Product: siehe CMP-202211001 - Anhang 1 - Produktübersicht
Datum FSN: (documentdatum) 06-12-2022

Waarschuwing van de fabrikant

Dit bericht is een waarschuwing van de fabrikant. Na het op de markt brengen van een medisch hulpmiddel is de fabrikant verplicht het hulpmiddel te volgen en waar nodig te verbeteren. Wanneer sprake is van een risico voor de gebruiker moet de fabrikant actie ondernemen. De fabrikant informeert gebruikers en de inspectie over deze actie met een waarschuwing (Field Safety Notice). Zie ook waarschuwingen medische hulpmiddelen. Dit hoort bij het toezicht op medische technologie.