Merit Medical Systems Inc, 1721504-08.01.22-004R, CorVocet Biopsy System
Merit Medical Systems, Inc. is voluntarily conducting a recall of the CorVocet™ Biopsy System due to a product defect resulting from incorrect assembly. This assembly error may occur when components are misaligned due to an issue with the manufacturing assembly equipment. In rare instances, the needle may eject from incorrectly assembled CorVocet™ biopsy devices when triggered outside the body, which results in the potential for a user or patient needle stick injury. The IFU includes the following precaution: “Never test the product by firing into the air. Damage may occur to the device and could result in patient and/or user injury.”
Fabrikant: Merit Medical Systems Inc.
Product: CorVocet Biopsy System
Datum FSN: 25-11-2022
Waarschuwing van de fabrikant
Dit bericht is een waarschuwing van de fabrikant. Na het op de markt brengen van een medisch hulpmiddel is de fabrikant verplicht het hulpmiddel te volgen en waar nodig te verbeteren. Wanneer sprake is van een risico voor de gebruiker moet de fabrikant actie ondernemen. De fabrikant informeert gebruikers en de inspectie over deze actie met een waarschuwing (Field Safety Notice). Zie ook waarschuwingen medische hulpmiddelen. Dit hoort bij het toezicht op medische technologie.