Gyrus ACMI, Inc. (Olympus Surgical Technologies America), QIL FY24-EMEA-26-FY24-OSTA-07, EZDilate Fixed Wire (Guided) Balloon

Olympus is taking this recall action after investigating reported complaints about issues inflating, deflating and/or retrieving the devices, as well as reports of bursting and leaking. Olympus has received reported adverse events involving foreign body in patient and prolonged procedure. Although our investigation is still ongoing, Olympus has learned that a misaligned processing aid during manufacturing may be a contributing factor. As a result, Olympus is requesting healthcare facilities return the affected model/lots to Olympus.
 

Fabrikant: Gyrus ACMI, Inc. (Olympus Surgical Technologies America)
Product: EZDilate Fixed Wire (Guided) Balloon
Datum FSN: 28 november 2023

Waarschuwing van fabrikant

Dit bericht is een waarschuwing van de fabrikant. Na het op de markt brengen van een medisch hulpmiddel is de fabrikant verplicht het hulpmiddel te volgen en waar nodig te verbeteren. Wanneer sprake is van een risico voor de gebruiker moet de fabrikant actie ondernemen. De fabrikant informeert gebruikers en de inspectie over deze actie met een waarschuwing (Field Safety Notice). Zie ook waarschuwingen medische hulpmiddelen. Dit hoort bij het toezicht op medische technologie.