Documenten - Waarschuwingen van fabrikanten medische hulpmiddelen

152 publicaties over Waarschuwingen van fabrikanten medische hulpmiddelen

Vul zoekcriteria in
Vul hier een trefwoord in waarop u wilt filteren.
Periode
U kunt hier een specifiek type publicatie kiezen, zoals brochures of jaarverslagen.
Sorteren op:
Abonneren op documenten over dit onderwerp
  1. Richard Wolf GmbH, FSN-FSCA700020652, FLEX. GRASP. FORCEPS or FLEX. BIOPSY FORCEPS

    We have received complaints from users stating that the forceps can no longer be opened during use. Our investigation has shown ...

    Waarschuwing | 23-03-2023

  2. Laborie Medical Technologies Corp., FSN-FSCA-2023-001, injeTAK Adjustable Tip Needle

    Deze brief heeft tot doel u te informeren over een vrijwillige terugroepactie voor de injeTAK instelbare naaldpunt (DIS199 en ...

    Waarschuwing | 13-03-2023

  3. Boston Scientific Corporation, FSN-92984351-FA, Habib EndoHPB Bipolar Radiofrequency Catheter

    Deze veiligheidskennisgeving levert belangrijke informatie met betrekking tot het gebruik van de Habib EndoHPB-katheter in ...

    Waarschuwing | 13-03-2023

  4. Philips Goldway Shenzhen Industrial Inc., 2022-CC-EC-010-IR, Efficia Pads Adapter Cable

    Dit probleem is vastgesteld als gevolg van een klacht van een klant. Philips heeft één melding ontvangen van een overlijden van ...

    Waarschuwing | 10-03-2023

  5. Zimmer Switzerland Manufacturing GmbH, ZFA2022-00302, CoCr Femoral Heads

    Zimmer GmbH is conducting a medical device Field Safety Corrective Action (correction) to update the compatibility matrix as ...

    Waarschuwing | 08-03-2023

  6. Intuitive Surgical, Inc, ISIFA2022-09-C, SureForm 45.SureForm 45 Curved-Tip.SureForm 60

    During standard post-market surveillance activities, Intuitive has observed an increase in complaints regarding engagement ...

    Waarschuwing | 09-02-2023

  7. Tobrix BV, 202203, radial fiber 600 mu

    Concerning LOT 2111049 of the 600mu radial fiber TXMF600R. Tip breaks of with little force. The broken tip can remain in the ...

    Waarschuwing | 01-12-2022

  8. Getinge Disinfection AB, FSN-CAPA 657266, Getinge ED-FLOW Range

    Getinge has discovered that washer disinfector ED-FLOW AER range potentially exhibit behavior due to a system error whereby, ...

    Waarschuwing | 03-11-2022

  9. Boston Scientific Corporation, FSN-92926415-FA, ORISE™ Gel Submucosal Lifting Agent

    Boston Scientific heeft onlangs ontdekt dat er bijwerkingen optraden die werden gekoppeld aan een reactie op vreemde voorwerpen ...

    Waarschuwing | 24-10-2022

  10. Coloplast A/S, FSN-FSCA_In-Ka_20221003, In-Ka Ureteral balloon dilatation catheter kit

    De vervaldatum op de IN-KA® ureterale ballondilatatie katheterset is niet correct. De houdbaarheid van één component in de set ...

    Waarschuwing | 13-10-2022