Homoth Medizinelektronik GmbH & Co.KG FSCA-LK-HO-15112019 - Luftkalorisator, Air Calorisator
As a part of our repair evaluation, Homoth Medizinelektronik has detected that, due to a possible production error, the medical devices li sted below {2012 to present year, blue LC displays) may not match with the relevant safety standard {DIN EN ISO 60601-1) in regards to fault safety. Potentially it is possible that an unfavorable defect of a component (short circuit of the circuit breaker to control the heating coil) may causes impermiss ible temperatures of the air flow and I or the handpiece. To our current knowledge the protective circuit designed for this case is not fully functional in all devices due to a faulty production. This event can Iead to impermissible temperatures (>+Sr C) with permanent activation of the heating coil.
Fabrikant: Homoth Medizinelektronik GmbH & Co.KG
Product: Luftkalorisator / Air Calorisator
Waarschuwing van de fabrikant
Dit bericht is een waarschuwing van de fabrikant. Na het op de markt brengen van een medisch hulpmiddel is de fabrikant verplicht het hulpmiddel te volgen en waar nodig te verbeteren. Wanneer sprake is van een risico voor de gebruiker moet de fabrikant actie ondernemen. De fabrikant informeert gebruikers en de inspectie over deze actie met een waarschuwing (Field Safety Notice). Zie ook waarschuwingen medische hulpmiddelen. Dit hoort bij het toezicht op medische technologie.