Abbott Molecular, FA-AM-NOV2022-283, Alinity m Resp-4-Plex AMP Kit
Abbott has received reports of an increase in reactive negative controls and false positive results, with certain lots of Alinity m Resp-4-PlexAMP Kit (CE), List Number 09N79-090 and of the Alinity m Resp-4-Plex AMP Kit (US), List Number 09N79-096 on the Alinity m System.
Specifically, an increase in the amount of reactive negative controls and false positive results have been reported for the influenza B virus (Flu B) and Respiratory Syncytial Virus (RSV) targets. Based on internal evaluation, false positive results and reactive negative controls manifest as a weak signa} with a late cycle number for these targets. Internal evaluation has not shown impact to influenza A virus (Flu A) and SARS targets.
Fabrikant: Abbott Molecular
Product: Alinity m Resp-4-Plex AMP Kit
Datum FSN: 22-11-2022
Waarschuwing van de fabrikant
Dit bericht is een waarschuwing van de fabrikant. Na het op de markt brengen van een medisch hulpmiddel is de fabrikant verplicht het hulpmiddel te volgen en waar nodig te verbeteren. Wanneer sprake is van een risico voor de gebruiker moet de fabrikant actie ondernemen. De fabrikant informeert gebruikers en de inspectie over deze actie met een waarschuwing (Field Safety Notice). Zie ook waarschuwingen medische hulpmiddelen. Dit hoort bij het toezicht op medische technologie.