intra special catheters GmbH FSCA200001 - PNEUMOCATH®_NEO-PNEUMOCATH®_THORACATH®-models
The background to the corrective action is an incident reported by a user concerning the product "NeoPneumocath for adults 10F - ...
-
Abena A/S FSN 1269554 - Abena Nova with MediSens Clip
Abena A/S experienced irregularities with the closing function of a clip. Hence, the clip was not tightening appropriately around ...
-
Turon MedTech AB D040307FSCA rev02 - CYTOGUIDE needle guide bushes
Some bushings have a material defect, which makes them brittle. The degree of this material defect varies. However, some of them, ...
-
Getinge Disinfection AB CAPA 219071 - Getinge 88-series
This is to inform you of a potential leakage that can occur on devices produced before November 2012. Our records indicate that ...
-
Luminex RC-19-012 - VERIGENE® External Barcode Scanner
An investigation has identified that the supplier of our VERIGENE External Barcode Scanner had issues scanning barcode labels ...
-
Tobrix BV, Tobrix radial fiber TXMF600R
Recentelijk zijn er een 3tal fibers met LOT nummer 1801005 defect gegaan gedurende de ELVA procedure.
-
FSN E-Drive - Decon Wheel AB
Decon Wheel AB has been informed that E-Drive can be mounted incorrectly on the Etac Cross 5 wheelchair. According to the ...
-
BioMérieux SA, FSCA #5763-1, BK VIRUS R-GENE
The aim of this communication is to inform you about low positive signals (aspecific signals below Ct corresponding to the limit ...
-
Covidien llc, FA1375, Hugo™ RAS Surgeon Console
The purpose of this letter is to advise you that Medtronic is initiating an Urgent Field Safety Notice for a specific subset of ...
-
Mikrogen GmbH, M23004, recomLine HIV Rev001 Build001.stc
Programming gap with an extremely rare band ratio of gp41 to gp36, resulting in a borderline output instead of positive, ...