Bio-Rad_DiaMed GmbH FSCA 004-19 - DiaClon Rh-Subgroups + K
Further to customer's reports, we have been able to confirm that interference can be observed in the anti-e microtube with QC ...
-
Becton Dickinson and Company, PAS-19-1454, BD Microtainer® Tubes
BD is overgegaan tot een corrigerende veiligheidsmaatregel om specifieke lots van de BD Microtainer® buizen uit omloop te nemen ...
-
Abbott Diagnostics FA10JUN2019 - Alinity ci-series System Control Module
Abbott has identified an issue with all on‐market versions of Alinity ci‐series Software where reuse of reaction vessels (RVs) ...
-
RSR Limited - RSR/06/2019/01 - RiaRSR™ AChRAb
This field safety notice is to inform you that the multiple lots of RiaRSR™ AChRAb may demonstrate high background counts.
-
DIAsource CAPA30680 - VIP RIA (Vasoactive intestinal polypeptide)
Low CPM values measured for the calibrators and kit controls generated with the kit lot 191601, which might lead to an unexpected ...
-
bioMérieux SA, FSCA 4379, VIDAS® NT-proBNP2
The investigation confirmed that the calibration of lot 1007078560 cannot be perfomed on VIDAS® PC instrument.
-
Hologic FA-00095 - Aptima Combo 2 Assay
Genetic variants are a natural by-product of microbial evolution and a new Chlamydia trachomatis (CT) variant has been identified ...
-
Roche Diagnostics GmbH - CN-432816 - cobas t 511, cobas t 711
The cobas t 511 and cobas t 711 coagulation analyzers have a waste management system that is based on weight sensors, which ...
-
Fast Track Diagnostics S.a.r.l. - FA-2019-05 - FTD Neuro 9
After internal investigation, Fast Track Diagnostics (FTD) has observed that nonspecific low positive signals may be generated ...
-
QIAGEN CI 3017461 - QIAsymphony Filter Tips, 1500 μl
Een klein percentage van de tips is beschadigd en er zijn schuurplekken aanwezig in het oppervlak van de tip. Deze schuurplekken ...