Documenten - Waarschuwingen van fabrikanten medische hulpmiddelen
584 publicaties over Waarschuwingen van fabrikanten medische hulpmiddelen
- Datum
- Relevantie
-
Hitachi High Technologies Corporation (HHT), FSN-CN-509982,CN-530429 (SBN-CPS- 2020-013), cobas 8000 Core Unit,cobas pro sample supply unit,cobas c 513 analyzer commercial system
Roche has received one customer complaint for the cobas 8000 modular analyzer series where due to a software limitation the ...
-
Roche Molecular Systems, FSN-CN-516475, cobas z 480 analyzer
Roche received complaint cases alleging false positive or invalid results with select cobas® 4800 IVD assays.
-
Siemens Healthcare Diagnostics Inc, FSN-ACHC 20-18, ADVIA and Atellica Fructosamine reagent
The purpose of this communication is to inform you of a positive bias with Quality Control (QC) and patient samples when using ...
-
Siemens Healthcare Diagnostics Inc, Inc, FSN-VC 20-03, Dimension Vista High Sensitivity Troponin I Flex
High Sensitivity Troponin I (TNIH) Flex® reagent cartridge Negative Bias with Patient Samples.
-
Roche Molecular Systems, FSN-CN-494142 (SB-RMD-2020-001), MagNA Pure 96 System
As a result of a customer complaint, it was discovered that, only when using the Sample Transfer protocol version 3.0 on the ...
-
Fast-Track Diagnostics S.a.r.l., FSN-FA-2020-04, FTD Stool parasites
Update van de gebruiksaanwijzing voor de FTD Stool parasites-kit.
-
Fast-Track Diagnostics S.a.r.l. FA-2020-02 FTD Vesicular rash
Het doel van deze communicatie is u te informeren over prestatieproblemen in verband met de ontoereikendheid van validatie en ...
-
Siemens IMC20-01 - IMMULITE 1000_2000 (XPi) Estradiol
The purpose of this communication is to inform you of high discordant Estradiol results for some patient samples with the ...
-
Siemens Healthcare Diagnostics Limited ASI 19-04 Atellica CH 930 Analyzer
The purpose of this communication is to inform you of issues with the Atellica CH 930 Analyzer listed in Table 1 above, installed ...
-
Roche Diagnostics GmbH CN-428741 (SBN-CPS-2019-014) ALB2 (Albumin Gen.2)BILT3 (Bilirubin Total Gen.3)
Roche has received a number of complaints about Albumin Gen.2 (ALB2) reagent lot 33962301 and Bilirubin Total Gen.3 (BILT3) ...