  1. Siemens Healthcare Diagnostics Products Ltd, IMC 24-02, IMMULITE® 2000 Anti-TG Ab

    Siemens Healthineers has confirmed, through the investigation of customer complaints, the potential for falsely elevated patient ...

    Waarschuwing | 25-06-2024

  2. Siemens Healthcare Diagnostics Inc., POC 24-015, RAPIDPoint® 500e Blood Gas System

    Siemens Healthineers has identified an issue with the RAPIDPoint 500e system software version 5.3. This issue affects how the ...

    Waarschuwing | 03-06-2024

  3. Agilent Technologies Singapore .International. Pte Ltd., Dako CoverStainer, Dako CoverStainer CPO

    It has come to our attention that the assembly supporting the weight of the lower front door in some cases may cause the ...

    Waarschuwing | 24-05-2024

  4. Alere San Diego,FCA-00000283, Cholestech LDX Analyzer

    This letter is to inform you that Alere San Diego has recently revised the Cholestech LDX™ Battery label warning to include the ...

    Waarschuwing | 23-05-2024

  5. Beckman Coulter, 24019, DxI 9000 Access Immunoassay Analyzer

    Beckman Coulter heeft vastgesteld dat de DxI 9000 Access Immunoassay-analysator kalibratorwaarden accepteert die zijn gescand ...

    Waarschuwing | 07-05-2024

  6. Stago, RC-24-0001, Star Max. Sta-R Evolution. Compact Max

    Stago has investigated and confirmed a potential risk of pipetting close to or into the packed red blood cells when using these ...

    Waarschuwing | 05-04-2024

  7. Abbott Laboratories, FA16JAN2024, Alinity s System

    Issue 1 Misassigned sample pipettor aspiration locations at the LAS can cause results to be wrongly attributed to the incorrect ...

    Waarschuwing | 25-01-2024

  8. Nova Biomedical Corporation, FCN 01-23, Stat Profile PRIME Plus

    lnitial start-up of a new BUN/Creatinine Sensor Cartridge may result in excessive sensor drift from the last automatic sensor ...

    Waarschuwing | 04-01-2024

  9. Becton, Dickinson and Company, IDS-23-4910, BD BACTEC™ MicroMGIT Calibration Vial

    Becton, Dickinson and Company has confirmed through 45 (forty-five) customer complaints that the BD BACTEC™ MicroMGIT™ ...

    Waarschuwing | 12-12-2023

  10. Actim Oy, 1698PARTUS, Actim PROM Test, Actim PROM 1ngeni Test

    Our recent interference testing has shown that the presence of lubricants in the vaginal specimen may interfere with the PROM ...

    Waarschuwing | 11-12-2023