  1. SENTINEL CH. S.p.A., 03/2023; FA02OCT2023, IRON, IRON, Alinity c Iron Reagent

    It has been identified that when using the ARCHITECT and Alinity c Iron products for the lots listed above, variable results can ...

    Waarschuwing | 12-10-2023

  2. Siemens Shenzhen Magnetic Resonance Ltd., MR078/23/S, Headband of the over-ear headphones

    This letter is to inform you about potential dot- or line-shaped artifacts caused by the presence of lubricating grease within ...

    Waarschuwing | 06-10-2023

  3. Medtronic Navigation, Inc., FA1361, Neuroscience CST - StealthStation™ S8 App version 2.0 and 2.0.1

    Medtronic heeft een softwareprobleem geïdentificeerd in de StealthStation™ S8-app, softwareversies 2.0 en 2.0.1 dat kan optreden ...

    Waarschuwing | 06-10-2023

  4. lmmunotech S.A.S. - A Beckman Coulter Company, 23043, Access hsTnI Reagent

    Beckman Coulter heeft vastgesteld dat er na de upgrade naar systeemsoftware 7.0.0 en hoger een restrisico op intratestoverdracht ...

    Waarschuwing | 05-10-2023

  5. Medeco B.V., FSN2023-01, Klinion Kliniray (Abdominal) Gauze Compress X-Ray

    After assessment of these products, it became known that x-ray threads in gauze compresses can disintegrate. It has been ...

    Waarschuwing | 05-10-2023

  6. TRUMPF Medizin Systeme GmbH Co. KG, FA-2023-047

    The potential impact for the missing IFU is that the users will not have awareness of appropriate use of the accessories, or ...

    Waarschuwing | 04-10-2023

  7. Covidien llc, FA1370, McGrath MAC Video Laryngoscope

    Medtronic is issuing this notification to inform the public of stolen defective  McGRATH™ MAC video laryngoscopes that have been ...

    Waarschuwing | 04-10-2023

  8. Seffiline S.R.L., SEFFIHAIR 3.1-SEFFICARE 3.1-SEFFIGYN 3.1-SEFFILLER 3.1 Seffiline S.r.l.

    During the assembly of the kits, two transfers for luerlock syringes supplied by a supplier other than the one qualified by ...

    Waarschuwing | 03-10-2023

  9. Atrium Medical Corporation, 3011175548-08.18.2023-001-C, Atrium Ocean Water Seal Chest Drains

    The Instructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution ...

    Waarschuwing | 03-10-2023

  10. Masimo Corporation, TC-I Reusable Tip Clip Sensor; MDSS ref: MDD23.148, TC-I Reusable Tip Clip Sensor

    Masimo identified TC- I Reusable Tip Clip sensors with higher than expected variability in SpO2 readings that are outside of the ...

    Waarschuwing | 03-10-2023