LeMaitre Vascular, CAPA 2022-001-NL, AlboGraft Polyester Vascular Graft
LeMaitre trekt bepaalde AlboGraft-apparaten terug en ruilt ze om vanwege een wijziging in de wettelijke status.
-
Baxter FA-2022-003 - TIP CAPS LUER SLIP YELLOW IND
Baxter Healthcare Corporation has received one complaint of a customer getting self-righting Luer lock tip caps, yellow (product ...
-
Mevion Medical Systems FSCA_220223 - Mevion S250i
An instance occurred at a customer site where a stuck motion button (see image below) on the Hand Pendant caused unintended Couch ...
-
Draeger Medical Systems PR116042 - Infinity® CentralStation SW VG2.1.2(SU1)
During the course of our global market surveillance activities, we have become aware of cases in which restarts of the lnfinity® ...
-
Siemens Healthineers FSCA AX002_20_S - Artis zee/Q/Q.zen Systems
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X- ray tube ...
-
Thermo Fisher Scientific, FSN-2022-002, ARIS HiQ
If an ESD pulse from a build-up of static electricity is discharged on the HiQ front door it can cause the instrument to go into ...
-
Liko AB, FA 2021-12-005-LUL-007, UNIVERSAL TWINBAR 670
Hillrom heeft vastgesteld dat er een mogelijkheid is dat patiënten kunnen vallen met de Universele Twinbar 670 QRH™, die mogelijk ...
-
STRYKER ENDOSCOPY, RA2022-2912780, 1688 Camera Control Unit (CCU)
A software defect has been identified in the 1688 Camera Control Unit (CCU) that will cause the image on the monitor to flip ...
-
Philips FSN-2021-IGT-BST-033 - Philips StentBoost Live R2.0 icm Philips Azurion SW R2.x
a) Als gevolg van een onjuiste configuratie-instelling in de EPX-database van StentBoost Live stopt de opname niet automatisch ...
-
NeuMoDx Molecular, a QIAGEN Company, FSN-CA-2021-006, NeuMoDx Cartridge
QIAGEN has become aware of the potential for false positive results to occur when the above lot of cartridges are used in ...