Documenten - Waarschuwingen van fabrikanten medische hulpmiddelen
584 publicaties over Waarschuwingen van fabrikanten medische hulpmiddelen
- Datum
- Relevantie
-
Instrumentation Laboratory_Werfen B.V. 2022-006-R - HemosIL ReadiPlasTin
Tot op heden hebben we 65 klachten (1 in Benelux) ontvangen van prestatieproblemen bij vier lotnummers HemosIL ReadiPlasTin. Deze ...
-
Siemens Healthcare Diagnostics AIMC 22-03 - Atellica® Solution and ADVIA Centaur® Systems
Siemens Healthcare Diagnostics Inc. has confirmed customer observations of falsely elevated results when using plasma specimens ...
-
Siemens Healthcare Diagnostics Inc, FSN-VC 22-01, Dimension Vista Magnesium Flex reagent cartridge
Siemens Healthineers has confirmed that the Dimension Vista Magnesium (MG) Flex reagent cartridge lots listed in Table 1 may show ...
-
Hologic FA-00197 - PANTHER FUSION EXTRACTION REAGENT-X kit
When the Panther Fusion Capture Reagent-X has been stored at 28oC or higher and is used to expiration, there is a possibility for ...
-
Bio-Rad_DiaMed GmbH FSCA 003-21 - ID-DiaCell I-II-III and ID-DiaCell IP-IIP-IIIP
Further to customer reports, we have been able to confirm that marked hemolysis can be observed with the ID-DiaCell I-II-III and ...
-
Abbott FA-AM-DEC2021-263 - Alinity m system
A software issue has been identified with the Alinity m System Software (Version 1.5.0 and greater) associated with the ability ...
-
B·R·A·H·M·S GmbH FSCA 02-21 - B·R·A·H·M·S KRYPTOR kits
Vials included only in the above listed B·R·A·H·M·S KRYPTOR kit lots might potentially leak during transport and when not stored ...
-
Fisher Diagnostics, Infinity Ammonia Reagent
Following a recent internal review of product technical documentation for the Infinity Ammonia Liquid Stable Reagent, Catalog # ...
-
bioMerieux SA, FSCA 5387, RAPID 20 E
Naar aanleiding van een klacht van een klant werd een fout in versie G van de RAPID 20 E™-bijsluiter vastgesteld.
-
Hain Lifescience GmbH, FSN-FSCA12102021, GenoType MTBDRplus V. 2.0 / GenoType MTBDRsl V.2.
We would like to inform you that the amplification control in the above-mentioned lots may show a weak positive signal in some ...