  1. PathoNostics B.V., AsperGenius Resistance Multiplex real-time PCR ki

    Negative results do not necessarily indicate absence of an (fungal) infection.

    Waarschuwing | 23-01-2024

  2. Gyrus ACMI, Inc., SG PK CUTTING FORCEPS, 5MM, 33CM. Olympus

    The jaw may fracture prior to the procedure during the inspection instructed per the IFU, or during the procedure.

    Waarschuwing | 23-01-2024

  3. Maquet Cardiopulmonary GmbH, 946521, H 31471#Hemoconcentrator Set BC 140 / H 15981#Set BC 140 Plus

    The component "Hemoconcentrator" was assembled into the finished product (Hemoconcentrator tubing set) after its shelf-life had ...

    Waarschuwing | 23-01-2024

  4. Gyrus ACMI, Inc., TFL Fiber Cleaver, TFL Fiber Stripper. Olympus

    Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a ...

    Waarschuwing | 23-01-2024

  5. OriGen Biomedical, Inc., FA-2023-001, Accessory Sets

    Deze kennisgeving betreft een terugroepactie/veiligheidsbericht met betrekking tot bepaalde partijen OriGen Biomedical® Accessory ...

    Waarschuwing | 22-01-2024

  6. Atrium Medical Corporation, 3011175548-12/13/2023-002-C, Drains Express Single

    On November 8, 2023, Getinge received notice from Nurse Assist, LLC that its Sterile Water, USP, 30mL syringes were being ...

    Waarschuwing | 22-01-2024

  7. GE Medical Systems SCS, FMI 34134, EVair medical air compressor, Evair 03

    GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are ...

    Waarschuwing | 22-01-2024

  8. B. Braun Melsungen AG, FSCA-VS-2023-01, Invasive blood pressure transducer set

    As users already know, prolonged contact with organic solvents (spray disinfectants) can promote the formation of stress cracks ...

    Waarschuwing | 22-01-2024

  9. Siemens Healthcare GmbH, AX010/23/S (FSCA), AX038/23/S (FSN), Sensis

    If the Sensis documentation functionality is used it may happen that during adding once-per-study reporting events (Type 1 ...

    Waarschuwing | 22-01-2024

  10. Philips Medical Systems Nederland BV, 2023-IGT-BST-020, Philips Azurion 7 M20 FlexArm system with Maquet M

    Philips is op de hoogte gebracht van twee mogelijke veiligheidsproblemen met het Azurion 7 M20 Flexarm-systeem ("Flexarm") met ...

    Waarschuwing | 22-01-2024