  1. Siemens Shanghai Medical Equipment Ltd., XP044/23/S, LUMINOS Impulse; Luminos RF Classic; Luminos Select; AXIOM Iconos MD; AXIOM Iconos R100

    This letter is to inform you about a prev entive measure to exclude any potential risk of patient injury caused by t he footrest ...

    Waarschuwing | 19-12-2023

  2. Philips Medical Systems Nederland BV, 2023-IGT-BST-024, Philips Azurion 7 M20 FlexArm system with Maquet M

    If, during a procedure, the detector is positioned at a distance of approximately four (4) cm from the patient, the connection ...

    Waarschuwing | 15-12-2023

  3. Philips Medical Systems Nederland BV, 2023-IGT-BST-005, Philips Azurion

    Philips has identified a potential issue which can cause the system to continuously restart (restart loop). This is likely to ...

    Waarschuwing | 15-12-2023

  4. GME German Medical Engineering GmbH, CAPA 20230928-5, FlexSys

    GME is currently preparing for certification within the new regulatory framework of the Medical Device Regulation (MOR). In case ...

    Waarschuwing | 15-12-2023

  5. Welch Allyn, Inc, FA-2023-055, Qstress 6, Xscribe 6

    Baxter Healthcare Corporation is issuing a Correction for the Q‐ Stress and XScribe Cardiac Stress Testing Systems (Q‐Stress ...

    Waarschuwing | 15-12-2023

  6. Maquet Critical Care AB, MCC-23-002-IU, Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e

    Internal tests at Getinge have revealed a cybersecurity related vulnerability which potentially could lead to Denial of Service ...

    Waarschuwing | 15-12-2023

  7. Ecolab Deutschland GmbH, ECL-FSCA-001, Incidin OxyWipe S; Incidin OxyFoam S

    Ecolab have retested the product’s efficacy of Incidin OxyWipe S and Incidin OxyFoam S against C. difficile according to the norm ...

    Waarschuwing | 15-12-2023

  8. BD, BDB-23-4854, BD® CD11b APC

    BD heeft aan de hand van acht (8) onderzoeken naar klachten van klanten bevestigd dat het bovenstaande lot van BD™ CD11b APC, CE ...

    Waarschuwing | 15-12-2023

  9. Cepheid, 23FAR014, Xpert Xpress CoV-2/Flu/RSV plus

    Bepaalde laboratoria die de Xpert® Xpress CoV-2/Flu/RSV plus-test en de beoogde monstertypen gebruiken, hebben een hoger aantal ...

    Waarschuwing | 15-12-2023

  10. Olympus Medical Systems Corporation, QIL FY24-EMEA-20-FY24-OMSC-20-GIF-1TH190, EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE

    As a result of testing, Olympus discovered EOG sterilization failures in which the GIF-1TH190 failed to achieve sterilization ...

    Waarschuwing | 15-12-2023